Cleanroom Management International
We care for your compliance
Contact

ATI Automatic Filter Tester TDA-100P

ATI Automatic Filter Tester TDA-100P
Description
The TDA-100P Automated Filter Tester utilizes state-of-the-art technology and microprocessor control. This design features a user-friendly interface to minimize setup and operator training. All functions and operating parameters of the TDA-100P can be programmed using just 4 function keys and the menus are shown on an easy-to read, alphanumeric, vacuum fluorescent (VF) display.

Features
The test parameters for Percent Penetration, Filter Resistance, and Filter Test Flow are shown on individual digital display panels. All three test parameters have individually adjustable high & low alarm set points which are continuously monitored by the microprocessor. Should a parameter exceed its set point, the FAIL light will illuminate and the parameter's digital display will flash. If operaing in Manual mode, testing will not be permitted until the alarm is reset by the operator.

The microprocessor self-checks the system during operation and the FAULT light alerts the operator to any system faults. The specific problem causing the fault is identified on the VF display. The near-monodispersed aerosol is also microprocessor controlled, allowing fully automatic, unattended operation when integrated into an automated production line.

The TDA-100P Automated Filter Tester is housed in a compact, tabletop cabinet which features an ergonomic front control panel. The TDA-100P has safety features on its test fixture that can be configured to accommodate any canister filter currently manufactured.
RS-232 and User Interface ports on the TDA-100P give the capability of data collection for SPC purposes and control of the machine by remote PC.

Specifications
AEROSOL GENERATION
Technique: Pneumatic nebulization using DOP/DEHP or PAO-4
Diameter: 0.185 µm CMD ± 0.020 µm (meets 42 CFR Part 84)
*0.32 µm approximate MMD (mass mean diameter)
Geometric Standard Deviation: <1.6 (meets 42 CFR Part 84)
Concentration: 20 to 150 milligrams per cubic meter (mg/m³)

AEROSOL DETECTION
Technique: Near forward light scattering
Dynamic Range: .0001 to 200 milligrams per cubic meter (mg/m³)
Accuracy: ± 1% of FS reading for the range in use
Sample Flow Rate: Full flow through detector, 10 to 100 standard liters per minute (slpm)
0.35 to 3.53 standard cubic feet per minute (standard cu ft/min)

FILTER TEST FLOW MEASUREMENT
Technique: NIST Traceable Mass Flow meter
Accuracy: ± 0.4% of reading + ± 0.2% of full scale
Operating Range*: 10 to 100 standard liters per minute (slpm) @ 21.1° C, 760 mm Hg
0.35 to 3.53 standard cubic feet per minute (scfm)

PRESSURE MEASUREMENT
Technique: Electronic pressure transducer
Accuracy: ± 1.25% of full scale
Operating Range*: 0 to 100 millimeters WC (0 to 3.9 inches WC)

EFFICIENCY MEASUREMENT
Flow Rate Through Media: 10 to 100 liters per minute (SLPM)
0.35 to 3.53 standard cubic feet per minute (standard cu ft/min)
Operating Range: Efficiencies to 99.9995%

OPERATIONAL REQUIREMENTS AND PHYSICAL CHARACTERISTICS
Power: 115 VAC, 60 Hertz, 2 Amps
Compressed Air: 283 SLPM @ 5.6 kg/cm² (10 SCFM @ 80 PSIG)
Size: 42" high × 22" wide × 30" deep (0.45 cubic meters)

*Note: Unit can be configured to customer requirements.

Benefits:
Faster Test Cycles of filter cartridges or media
Automated control of testing conditions
Minimal requirement for operator time or attention
Eliminates operator error
More reliable operation and test results
Less downtime and operator adjustments
Obtain SPC data
UID: 1.08.009