Cleanroom Management International
We care for your compliance
Contact

Ecolog

ECOLOG TN

ECOLOG TN are universal, multi-purpose instruments for monitoring and documenting temperatures. The possible applications for thes measuring instruments are practically limitless, ranging from monitoring rooms or production processes to documenting transport conditions. Versatile configuration possibilities, sensors for various applications and a comprehensive assortment of accessories ensure reliable solutions for every requirement.

  • 2 - 4 channel dataloggers with internal and external sensors.
  • Principle of measurement : resistance measuring (NTC).
  • Large display for current measured values and alarm information.
  • Inputs for recording digital states.
  • Evaluation with elproLOG ANALYZE software at PC.
  • Programmable recording interval.
  • One year memory capacity for 64.000 measured values.
  • Keypad on datalogger (TN4, TN4-L) :
    • Display measured values and alarms.
    • Controls printing functions.
    • Tests alarm output.
    • Resets alarm output.
    • Starts logging.
  • Open interface for communicationwith other devices (TN3-P, TN4, TN4-L).
  • Low energy consumption due to power-save mode.
  • Also available with EEx approval.

 

 

ECOLOG TH

For users who set high standards, the ECOLOG TH is the ideal instrument for logging temperature and relative humidity. High reliability and a proven measuring technique, versatile configuration possibilities and the comprehensive assortment of accessories are some of the reasons for the succes of the ECOLOG TH in a multitude of industries and applications.

  • Temperature and relative humidity logging-with internal and external sensors.
  • Principle of measurement : T: resistance measuring (NTC), Hr: capacitive.
  • Implementation of calibrated and replaceable combi-sensors.
  • Large display for current measured values and alarm information.
  • Inputs for recording digital states.
  • Evaluation with elproLOG ANALYZE at PC.
  • Programmable recording interval.
  • One-year memory capacity for 64.000 measured values.
  • Keypad at datalogger :
    • Displays measured values and alarms.
    • Controls printing functions.
    • Tests alarm output.
    • Resets alarm output.
    • Starts logging.
    • Measured value marking.
  • Open interface for communication with other devices.
  • Low energy consumption due to power-save mode.
  • Also available with EEx approval.

ECOLOG TP

The ECOLOG TP stands for precision measuring and evaluating of temperatures in the -200°C .. 550°C range. This universal datalogger, based on the standardized PT100 measuring technique and with a comprehensive assortment of sensors and accessories, can be used for widespread applications ranging from monitoring low temperature to documenting drying processes.

  • 2 - 4 channel datalogger.
  • Principle of measurement : 4-wire PT100 in accordance with IEC751 / EN60751.
  • Digital inputs and evaluation with elproLOG ANALYZE software at PC.
  • Large display for current measured values and alarm information.
  • Programmable recording interval.
  • One-year memory capacity for 64.000 values.
  • Keypad at datalogger :
    • Displays measured values and alarms.
    • Controls printing functions.
    • Resets alarm output.
    • Starts logging.
  • Open interface for communication with other devices.
  • Low energy consumption due to power-save mode.
  • Also available with EEx approval.

 

 

Temperature measurement TPx (only datalogger at room temperature)

 Measurement range          Resolution        Linearity

 -200°C .. -101°C                   0.2°C             ±0.3°C

 -100°C .. 399°C                    0.1°C             ±0.2°C

   400°C .. 499°C                    0.1°C             ±0.3°C

   500°C .. 550°C                    0.2°C             ±0.5°C

  

Accuracy of measurement rel. humidity

Précision (@ 23°C)       Hystérésis (10-90-10% rH)

     ±1.5% rH                       

  

Temperature measurement TNx & THx (Datalogger with temperature sensor, logger at room temperature)

Measurement range        Resolution               Typical accuracy (U95)

   -50°C .. -31°C                 0.1°C                          ±0.6°C

   -30°C .. -16°C                 0.1°C                          ±0.4°C

   -15°C .. -1°C                   0.1°C                          ±0.3°C

      0°C .. 49°C                  0.1°C                           ±0.2°C

    50°C .. 99°C                   0.1°C                          ±0.3°C

   100°C .. 125°C                0.1°C                           ±0.6°C

   126°C .. 140°C                0.1°C                           ±0.9°C

 

 

Products

ECOLOG TN2 Temperature Datalogger
ECOLOG TN3-P Temperature Datalogger
ECOLOG TN4 Temperature Datalogger
ECOLOG TN4-L Temperature Datalogger
ECOLOG TH1 Temperature and Humidity Datalogger
ECOLOG TP4-L Temperature Datalogger
ECOLOG TP2 Temperature Datalogger
ECOLOG TH2 Temperature and Humidity Datalogger
USB to RS232 converter, CD and manual, direct connection to PC USB-port (COM1-COM6)
PC data cable ECOLOG, HOTBOX-xx
Vorige   1 2   Volgende